These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Chiral recognition of mandelic acid by L-phenylalanine-modified sensor using quartz crystal microbalance.
    Author: Guo HS, Kim JM, Chang SM, Kim WS.
    Journal: Biosens Bioelectron; 2009 May 15; 24(9):2931-4. PubMed ID: 19282168.
    Abstract:
    This study presents a new method for the highly selective recognition of chiral mandelic acid (MA) using L-phenylalanine (L-Phe) as the selector. The proposed method is based on quartz crystal microbalance (QCM) detection, integrated with a vapor diffused molecular assembly (VDMA) reaction technique. The construction of the L-Phe-modified QCM sensor involved a two-step assembly procedure. The chiral recognizability of L- and D-MA on the L-Phe-modified surface was then examined using the VDMA method and QCM. The chiral discrimination factor between L- and D-MA detected by QCM, alpha(L-MA/D-MA), was found to be about 8. The VDMA selective sensing of L-MA on the L-Phe-modified surface was also confirmed by the contact angle measurements. The L-Phe-modified QCM sensor showed good stability and reusability. The present chiral recognition results suggest that L-phe is an excellent resolving agent for the resolution of chiral mandelic acid.
    [Abstract] [Full Text] [Related] [New Search]