These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Syringe shape and positioning relative to efficiency volume inside dose calibrators and its role in nuclear medicine quality assurance programs. Author: Santos JA, Carrasco MF, Lencart J, Bastos AL. Journal: Appl Radiat Isot; 2009 Jun; 67(6):1104-9. PubMed ID: 19282196. Abstract: A careful analysis of geometry and source positioning influence in the activity measurement outcome of a nuclear medicine dose calibrator is presented for (99m)Tc. The implementation of a quasi-point source apparent activity curve measurement is proposed for an accurate correction of the activity inside several syringes, and compared with a theoretical geometric efficiency model. Additionally, new geometrical parameters are proposed to test and verify the correct positioning of the syringes as part of acceptance testing and quality control procedures.[Abstract] [Full Text] [Related] [New Search]