  • Title: A silicon-based, three-dimensional neural interface: manufacturing processes for an intracortical electrode array.
    Author: Campbell PK, Jones KE, Huber RJ, Horch KW, Normann RA.
    Journal: IEEE Trans Biomed Eng; 1991 Aug; 38(8):758-68. PubMed ID: 1937509.
    Abstract:
    A method has been developed for the manufacture of a "three-dimensional" electrode array geometry for chronic intracortical stimulation. This silicon based array consists of a 4.2 x 4.2 x 0.12 mm thick monocrystalline substrate, from which project 100 conductive, silicon needles sharpened to facilitate cortical penetration. Each needle is electrically isolated from the other needles, and is about 0.09 mm thick at its base and 1.5 mm long. The sharpened end of each needle is coated with platinum to facilitate charge transfer into neural tissue. The following manufacturing processes were used to create this array. 1) Thermomigration of 100 aluminum pads through an n-type silicon block. This creates trails of highly conductive p+ silicon isolated from each other by opposing pn junctions. 2) A combination of mechanical and chemical micromachining which creates individual penetrating needles of the p+ silicon trails. 3) Metal deposition to create active electrode areas and electrical contact pads. 4) Array encapsulation with polyimide. The geometrical, mechanical, and electrical properties of these arrays should make them well suited as interfaces to cortical tissue.