These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Development of a minimally invasive implantable wireless vital signs sensor platform. Author: Tan R, McClure T, Schulam P, Schmidt J. Journal: Stud Health Technol Inform; 2009; 142():380-5. PubMed ID: 19377188. Abstract: A wireless implantable pressure sensing system has been developed for short term urological studies. In in vitro testing, it was able to sense a physiologically relevant pressure range (0-1.5 psi gauge) with a resolution of 0.02 psi. In vivo testing demonstrated its ability to operate for >90 hours while implanted in a porcine model. Possible urination events and peristalsis were measured. The design of a second generation of this device is also presented.[Abstract] [Full Text] [Related] [New Search]