These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: A simple automated solid-phase extraction procedure for measurement of 25-hydroxyvitamin D3 and D2 by liquid chromatography-tandem mass spectrometry.
    Author: Knox S, Harris J, Calton L, Wallace AM.
    Journal: Ann Clin Biochem; 2009 May; 46(Pt 3):226-30. PubMed ID: 19389886.
    Abstract:
    BACKGROUND: Measurement of 25-hydroxyvitamin D(3) (25OHD(3)) and D(2) (25OHD(2)) is challenging. Liquid chromatography-tandem mass spectrometry (LC-MS/MS) methods have been described but they are often complex and difficult to automate. We have developed a simplified procedure involving an automated solid-phase extraction (SPE). METHODS: Internal standard (hexadeuterated 25-hydroxyvitamin D(3)) was added to serum or plasma followed by protein precipitation with methanol. Following centrifugation, a robotic instrument (CTC PAL [Presearch] for ITSP SPE [MicroLiter Analytical Supplies, Inc]) performed a six-step SPE procedure and the purified samples were injected into the LC-MS/MS. Quantification of 25OHD(3) and 25OHD(2) was by electrospray ionization MS/MS in the multiple-reaction monitoring mode. RESULTS: The lower limit of quantitation was 4.0 nmol/L for 25OHD(3) and 7.5 nmol/L for 25OHD(2). Within- and between-assay precision was below 10% over the concentration range of 22.5-120 nmol/L for D(3) and 17.5-70 nmol/L for D(2) (n = 10). The calibration was linear up to 2500 nmol/L (r = 0.99). Recoveries ranged between 89% and 104% for both metabolites and no ion suppression was observed. The results obtained compared well (r = 0.96) with the IDS-OCTEIA 25-hydroxyvitamin D enzyme immunoassay for samples containing less than 125 nmol/L, at higher concentrations the immunodiagnostic system (IDS) method showed positive bias. CONCLUSIONS: Our simplified sample preparation and automated SPE method is suitable for the measurement of 25OHD(3) and D(2) in a routine laboratory environment. The system can process up to 300 samples per day with no cumbersome solvent evaporation step and minimal operator intervention.
    [Abstract] [Full Text] [Related] [New Search]