These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Determination of naltrexone and 6 beta-naltrexol in plasma by high-performance liquid chromatography with coulometric detection.
    Author: Zuccaro P, Altieri I, Betto P, Pacifici R, Ricciarello G, Pini LA, Sternieri E, Pichini S.
    Journal: J Chromatogr; 1991 Jul 05; 567(2):485-90. PubMed ID: 1939481.
    Abstract:
    A simple and reliable reversed-phase high-performance liquid chromatographic method with coulometric detection is described for the quantitation of naltrexone and its metabolite, 6 beta-naltrexol, in plasma samples of healthy volunteers who received orally 50 mg of naltrexone. The analytes and the internal standard, naloxone, are extracted with an octadecyl solid-phase extraction column before chromatography. The mobile phase is 0.01 M potassium phosphate (pH 3)-acetonitrile (85:15, v/v) and it is pumped at 0.8 ml/min. The coulometric detector is formed by two electrodes set at +0.20 V and +0.70 V, with a palladium reference electrode. The limit of quantitation observed was 5 ng/ml for both naltrexone and 6 beta-naltrexol. This method can be used to investigate pharmacokinetic parameters of different pharmaceutical preparations of this opioid antagonist.
    [Abstract] [Full Text] [Related] [New Search]