These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Development of microfluidic chips for heterogeneous receptor-ligand interaction studies.
    Author: Goldberg MD, Lo RC, Abele S, Macka M, Gomez FA.
    Journal: Anal Chem; 2009 Jun 15; 81(12):5095-8. PubMed ID: 19441833.
    Abstract:
    A simple microfluidic-based technique to quantitate the binding affinity between the glycopeptide antibiotics teicoplanin from Actinoplanes teicomyceticus and vancomycin from Streptomyces orientalis and 5-carboxyfluorescein-D-Ala-D-Ala-D-Ala (5-FAM-(DA)(3)) is described. In this work, (3-aminopropyl)triethoxysilane is used to modify the surfaces of a series of microchannels, and each channel is subsequently exposed to a solution of antibiotic for a few minutes. The antibiotic is retained after washing through electrostatic interactions, and the series of channels are subsequently exposed to an increasing concentration of 5-FAM-(DA)(3) followed by washing to exclude any nonspecific binding. The extent of fluorescence is quantified using a microscope fitted with a CCD camera. The binding constants for the interaction of teicoplanin and vancomycin with the fluorescent peptide were determined to be 6.03 +/- 0.97 x 10(4) and 4.93 +/- 1.13 x 10(4) M(-1), respectively, in good agreement with previous data. The ease of quantifying the extent of interaction in this microchip technique may prove powerful for exploration of a myriad of receptor-ligand pairs.
    [Abstract] [Full Text] [Related] [New Search]