  • Title: Development of a new measurement method for serum calcium with chlorophosphonazo-III.
    Author: Hokazono E, Osawa S, Nakano T, Kawamoto Y, Oguchi Y, Hotta T, Kayamori Y, Kang D, Cho Y, Shiba K, Sato K.
    Journal: Ann Clin Biochem; 2009 Jul; 46(Pt 4):296-301. PubMed ID: 19454535.
    Abstract:
    BACKGROUND: Although serum calcium has been measured using the o-cresolphthalein complexone (oCPC) method in the clinical laboratory, this method still has some problems regarding linearity and reagent stability. We developed a new measurement procedure using chlorophosphonazo-III (CPZ-III: 2,7-bis (4-chloro-2-phosphonophenylazo) -1,8- dihydroxy-3, 6-naphthalenedisulphonic acid, disodium salt) as a chelator with an acid medium for serum calcium measurement. The present method showed better linearity and reagent stability compared with the oCPC method. METHODS: Characteristics were studied in optimized conditions measuring wavelength by absorption spectra analysis, and interference of protein and metals with Mg(2+), Fe(2+), Cu(2+) and Zn(2+). The method was applied to an automated analyser (7170; Hitachi High Technologies Corp). The measurement performance was evaluated for accuracy, precision, recovery rate, linearity and reagent stability with a comparison study against atomic absorption spectrophotometry (AAS). RESULTS: The within-run and between-run variations (coefficient of variation [CV]) were 0.92-1.01% and 0.75-1.43%, respectively. The linearity was 0-7.0 mmol/L. The comparison study obtained y = 1.002x (AAS) - 0.10, Sy/x = 0.18 mmol/L, n = 50. Reagent stability was at least 20 d at 4 degrees C without daily calibration. CONCLUSION: The new calcium measurement method in serum was demonstrated to have reliable and acceptable performances as a routine test in clinical laboratory.
    [Abstract] [Full Text] [Related] [New Search]