  • Title: A highly sensitive and robust UPLC-MS with electrospray ionization method for quantitation of taxifolin in rat plasma.
    Author: Wang X, Xia H, Xing F, Deng G, Shen Q, Zeng S.
    Journal: J Chromatogr B Analyt Technol Biomed Life Sci; 2009 Jun 15; 877(18-19):1778-86. PubMed ID: 19464238.
    Abstract:
    A sensitive ultra performance liquid chromatography-mass spectrometry method has been developed and validated for the quantification of taxifolin in rat plasma. Following liquid/liquid extraction by ethyl acetate, the analytes were separated on a Sunfire (2.1 mm x 50 mm, 3.5 microm) column and analyzed in the selected ion recording with a negative electrospray ionization mode. The method was linear over the concentration range of 6-6750 ng/mL. Intra- and inter-day precisions were all within 8% and accuracy ranged from 92.9% to 105.1%. The lower limit of quantification was 6 ng/mL. The present method was successfully applied to the estimation of the pharmacokinetic parameters of taxifolin following intravenous and oral administration to rats. The absolute bioavailability of taxifolin was 0.17% in rat.