  • Title: The influence of a Si cantilever tip with/without tungsten coating on noncontact atomic force microscopy imaging of a Ge(001) surface.
    Author: Naitoh Y, Kinoshita Y, Jun Li Y, Kageshima M, Sugawara Y.
    Journal: Nanotechnology; 2009 Jul 01; 20(26):264011. PubMed ID: 19509444.
    Abstract:
    A sharp probe tip with atomic scale stability is essential and desirable for noncontact atomic force microscopy (NC-AFM) studies at the atomic scale. We observed a Ge(001) surface using both a Si cantilever and a tungsten coated Si cantilever at room temperature in order to investigate the influence of the tip apex structure on the NC-AFM images. By using the Si cantilever, we first obtained four types of image at the atomic scale which can be explained assuming a dimer structure on the tip apex. On the other hand, the home-made tungsten coated tip, which has atomic scale stability and high electric conductivity, imaged the so-called ordered c(4 x 2) structure without any artifacts. The tungsten coated cantilever was found to have significantly higher performance for NC-AFM studies at the atomic scale than the Si cantilever.