These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: The influence of a Si cantilever tip with/without tungsten coating on noncontact atomic force microscopy imaging of a Ge(001) surface. Author: Naitoh Y, Kinoshita Y, Jun Li Y, Kageshima M, Sugawara Y. Journal: Nanotechnology; 2009 Jul 01; 20(26):264011. PubMed ID: 19509444. Abstract: A sharp probe tip with atomic scale stability is essential and desirable for noncontact atomic force microscopy (NC-AFM) studies at the atomic scale. We observed a Ge(001) surface using both a Si cantilever and a tungsten coated Si cantilever at room temperature in order to investigate the influence of the tip apex structure on the NC-AFM images. By using the Si cantilever, we first obtained four types of image at the atomic scale which can be explained assuming a dimer structure on the tip apex. On the other hand, the home-made tungsten coated tip, which has atomic scale stability and high electric conductivity, imaged the so-called ordered c(4 x 2) structure without any artifacts. The tungsten coated cantilever was found to have significantly higher performance for NC-AFM studies at the atomic scale than the Si cantilever.[Abstract] [Full Text] [Related] [New Search]