These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Ultra-high-speed optical coherence tomography with a stretched pulse supercontinuum source.
    Author: Moon S, Kim DY.
    Journal: Opt Express; 2006 Nov 27; 14(24):11575-84. PubMed ID: 19529577.
    Abstract:
    We introduce a new high-speed Fourier-domain optical coherence tomography (FD-OCT) scheme based on a stretched pulse supercontinuum source. A wide-band short pulse of a supercontinuum source of which output spectrum spanned a wavelength range from 1,200 nm to 1,550 nm was stretched to a long pulse of 70-ns duration by using a dispersive fiber due to the group-velocity dispersion, and it was used directly as frequency-swept light for FD-OCT. The OCT spectral interferogram was acquired in the time domain and converted into the spectral domain by the pre-calibrated time-to-wavelength relation. Using this stretched-pulse OCT (SP-OCT) scheme, we have demonstrated an ultrahigh-speed axial-line scanning rate of 5 MHz. The axial resolution of 8 microm was achieved without re-calibration of the sweep characteristic owing to the passive nature of the frequency-sweeping mechanism.
    [Abstract] [Full Text] [Related] [New Search]