  • Title: A system of calibrating microtomography for use in caries research.
    Author: Schwass DR, Swain MV, Purton DG, Leichter JW.
    Journal: Caries Res; 2009; 43(4):314-21. PubMed ID: 19556791.
    Abstract:
    Desktop microcomputed tomography (micro-CT) offers a non-invasive 3-dimensional analysis of structures and their physical properties. To date, the use of micro-CT has mostly involved qualitative observations, with the extent of quantitative analysis relying on automated internal calibration by the micro-CT control software. However, the value of such calibration is limited by machine drift. For an accurate quantitative use of micro-CT, it is recognized that external means of calibration are needed. A novel system of calibration standards, also known as 'phantoms', is presented. A range of low mineral concentration phantoms involving triethylene glycol dimethacrylate/glycerolate dimethacrylate resin mixed with commercial pure hydroxyapatite (HAP), from 0.07 to 1.05 g/cm(3), was fabricated. Sintered HAP was impregnated with the same resin, producing phantoms with medium-level mineral concentrations up to 1.90 g/cm(3). These phantoms were easy to create, proved accurate and stable with repeated use, and were found to mimic the composite nature of dental enamel and dentine structures under investigation.