These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: A system of calibrating microtomography for use in caries research. Author: Schwass DR, Swain MV, Purton DG, Leichter JW. Journal: Caries Res; 2009; 43(4):314-21. PubMed ID: 19556791. Abstract: Desktop microcomputed tomography (micro-CT) offers a non-invasive 3-dimensional analysis of structures and their physical properties. To date, the use of micro-CT has mostly involved qualitative observations, with the extent of quantitative analysis relying on automated internal calibration by the micro-CT control software. However, the value of such calibration is limited by machine drift. For an accurate quantitative use of micro-CT, it is recognized that external means of calibration are needed. A novel system of calibration standards, also known as 'phantoms', is presented. A range of low mineral concentration phantoms involving triethylene glycol dimethacrylate/glycerolate dimethacrylate resin mixed with commercial pure hydroxyapatite (HAP), from 0.07 to 1.05 g/cm(3), was fabricated. Sintered HAP was impregnated with the same resin, producing phantoms with medium-level mineral concentrations up to 1.90 g/cm(3). These phantoms were easy to create, proved accurate and stable with repeated use, and were found to mimic the composite nature of dental enamel and dentine structures under investigation.[Abstract] [Full Text] [Related] [New Search]