These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Simple and rapid voltammetric determination of morphine at electrochemically pretreated glassy carbon electrodes. Author: Li F, Song J, Gao D, Zhang Q, Han D, Niu L. Journal: Talanta; 2009 Aug 15; 79(3):845-50. PubMed ID: 19576454. Abstract: A simple and rapid method for morphine detection has been described based on electrochemical pretreatment of glassy carbon electrode (GCE) which was treated by anodic oxidation at 1.75 V, following potential cycling in the potential range from 0 V to 1.0 V vs. Ag|AgCl reference electrode. The sensitivity for morphine detection was improved greatly and the detection limit was 0.2 microM. The reproducibility of the voltammetric measurements was usually less than 3% RSD for six replicate measurements. Moreover, this method could readily discriminate morphine from codeine. And an electrochemical detection of morphine in spiked urine sample was succeeded with satisfactory results.[Abstract] [Full Text] [Related] [New Search]