Search MEDLINE/PubMed
Title: Susceptibility effects in hyperpolarized (3)He lung MRI at 1.5T and 3T. Author: Deppe MH, Parra-Robles J, Ajraoui S, Parnell SR, Clemence M, Schulte RF, Wild JM. Journal: J Magn Reson Imaging; 2009 Aug; 30(2):418-23. PubMed ID: 19629971. Abstract: PURPOSE: To compare susceptibility effects in hyperpolarized (3)He lung MRI at the clinically relevant field strengths of 1.5T and 3T. MATERIALS AND METHODS: Susceptibility-related B(0) inhomogeneity was evaluated on a macroscopic scale by B(0) field mapping via phase difference. Subpixel susceptibility effects were quantified by mapping T2. Comparison was made between ventilation images obtained from the same volunteers at both field strengths. RESULTS: The B(0) maps at 3T show enhanced off-resonance effects close to the diaphragm and the ribs due to susceptibility differences. The average T2 from a voxel (20 x 4 x 4) mm(3) was determined as T2 = 27.8 msec +/- 1.2 msec at 1.5T compared to T2 = 14.4 msec +/- 2.6 msec at 3T. In ventilation images the most prominent effect is increased signal attenuation close to the intrapulmonary blood vessels at higher B(0). CONCLUSION: Image homogeneity and T2 are lower at 3T due to increased B(0) inhomogeneity as a consequence of susceptibility differences. These findings indicate that (3)He imaging at 3T has no obvious benefit over imaging at 1.5T, as signal-to-noise ratio (SNR) was comparable for both fields in this work.[Abstract] [Full Text] [Related] [New Search]