These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: An evaluation of automated methods for measurement of serum 25-hydroxyvitamin D.
    Author: Wagner D, Hanwell HE, Vieth R.
    Journal: Clin Biochem; 2009 Oct; 42(15):1549-56. PubMed ID: 19631201.
    Abstract:
    OBJECTIVES: To compare two new automated assays with the well-established reference method, DiaSorin radioimmunoassay (RIA), for quantitation of serum total 25-hydroxyvitamin D [25(OH)D]. METHODS: 25(OH)D from human sera (n=158) was measured using DiaSorin RIA and two automated platforms, DiaSorin "LIAISON 25 OH Vitamin D TOTAL", and Roche Modular "Vitamin D3 (25-OH)". Methods were compared by regression and Bland-Altman analyses. RESULTS: DiaSorin LIAISON demonstrated a stronger correlation (r=0.918) and better agreement (bias=-0.88 nmol/L) with DiaSorin RIA than the Roche Modular assay (r=0.871, bias=-2.55 nmol/L). Precision ranges (CV%) for the RIA, LIAISON, and Roche Modular assays, respectively, were: within run (6.8-12.9%, 2.8-8.1%, and 1.9-5.5%), and total precision (7.4-14.5%, 7.3-17.5%, and 7.6-14.5%). CONCLUSION: DiaSorin LIAISON displayed the best correlation and agreement with DiaSorin RIA. The DiaSorin LIAISON 25 OH Vitamin D TOTAL assay is an accurate and precise automated tool for serum total 25(OH)D determination.
    [Abstract] [Full Text] [Related] [New Search]