These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Application of real-time three-dimensional transesophageal echocardiography using a matrix array probe for transcatheter closure of atrial septal defect.
    Author: Taniguchi M, Akagi T, Watanabe N, Okamoto Y, Nakagawa K, Kijima Y, Toh N, Ohtsuki S, Kusano K, Sano S.
    Journal: J Am Soc Echocardiogr; 2009 Oct; 22(10):1114-20. PubMed ID: 19647397.
    Abstract:
    BACKGROUND: The aim of this study was to demonstrate the utility of real-time three-dimensional (3D) transesophageal echocardiography (RT3D-TEE) using a matrix array 3D transesophageal echocardiographic probe for morphologic evaluation and guidance of transcatheter closure of atrial septal defects (ASDs). METHODS: Forty-eight consecutive patients scheduled for the intervention were included. Two-dimensional (2D) transesophageal echocardiography (2D-TEE) and RT3D-TEE were performed before and during the procedures. Measurements of maximal ASD diameter and surrounding rims obtained on RT3D-TEE were compared with those obtained on 2D-TEE. RESULTS: In 46 patients (96%), optimal 3D images for the morphologic evaluation of ASDs were obtained. RT3D-TEE facilitated the evaluation of ASD morphology and surrounding rims and was able to provide intraprocedural information clearly. A Bland-Altman plot showed a mean maximal diameter difference of -0.12 mm between the means (95% limits of agreement, -2.2 to 2.5 mm). CONCLUSION: RT3D-TEE is a clinically useful, complementary option to 2D-TEE for evaluation of ASD morphology and for interventional guidance.
    [Abstract] [Full Text] [Related] [New Search]