These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Impact of stem-broach sizing on the cement mantle of Lubinus SP II stems. A CT scan analysis. Author: Scheerlinck T, de Mey J, Deklerck R. Journal: Acta Orthop Belg; 2009 Jun; 75(3):346-53. PubMed ID: 19681321. Abstract: Lubinus SP II stems are cemented either line-to-line with the largest broach or one-size undersized. The purpose of this study was to compare both implantation techniques. We used 18 polymeric stem replicas cemented line-to-line and undersized in paired cadaveric femora and analyzed them with CT scan images. Cementing Lubinus stems line-to-line resulted in higher medullary canal-filling indices (28.26 +/- 4.10%), thinner cement mantles (3.29 +/- 0.40 mm), more cement defects (5.12 +/- 1.69%) and more areas of thin cement (23.81 +/- 7.13%) than undersizing (respectively: 23.61 +/- 4.24%, 3.62 +/- 0.43 mm, 1.48 +/- 2.04%, 15.11 +/- 5.93%). In both settings, over 80% of areas of thin or deficient cement were supported by cortex. Using a line-to-line technique, adequate stem alignment was achieved without distal centralizer. Undersizing the stem and using a distal centralizer reduced the incidence of distal cement defects by a factor 10. While stems cemented line-to-line might have mechanical advantages, undersizing and using distal centralizers reduced potential pathways for debris migration to the bone-cement interface.[Abstract] [Full Text] [Related] [New Search]