These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: White-light scanning interferometer for absolute nano-scale gap thickness measurement. Author: Xu Z, Shilpiekandula V, Youcef-toumi K, Yoon SF. Journal: Opt Express; 2009 Aug 17; 17(17):15104-17. PubMed ID: 19687988. Abstract: A special configuration of white-light scanning interferometer is described for measuring the absolute air gap thickness between two planar plates brought into close proximity. The measured gap is not located in any interference arm of the interferometer, but acts as an amplitude-and-phase modulator of the light source. Compared with the common white-light interferometer our approach avoids the influence of the chromatic dispersion of the planar plates on the gap thickness quantification. It covers a large measurement range of from approximate contact to tens of microns with a high resolution of 0.1 nm. Detailed analytical models are presented and signal-processing algorithms based on convolution and correlation techniques are developed. Practical measurements are carried out and the experimental results match well with the analysis and simulation. Short-time and long-time repeatabilities are both tested to prove the high performance of our method.[Abstract] [Full Text] [Related] [New Search]