These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: A review of reprocessing techniques of flexible nasopharyngoscopes.
    Author: Collins WO.
    Journal: Otolaryngol Head Neck Surg; 2009 Sep; 141(3):307-10. PubMed ID: 19716004.
    Abstract:
    To provide assistance to otolaryngologists to decide the best manner in which to reprocess flexible nasopharyngoscopes, a review of existing English language medical literature regarding the methods of flexible endoscope reprocessing was performed, including previously published guidelines from other medical disciplines. Multiple steps were confirmed to be critical to effectively reprocess flexible nasopharyngoscopes. High-level disinfection has been determined to be the minimum level of disinfection required for reprocessing of flexible nasopharyngoscopes. Several steps are important in all reprocessing techniques, including manual cleaning, leak testing, cleaning with an enzymatic agent, high-level disinfection, and drying with vertical storage. Three techniques are available to achieve high-level disinfection: manual disinfection with a liquid disinfectant/sterilant, use of an automated endoscope reprocessor, and use of a disposable sheath. Achieving high-level disinfection of flexible nasopharyngoscopes can be accomplished by a variety of methods. Strict adherence to recommended procedures is critical.
    [Abstract] [Full Text] [Related] [New Search]