These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: A novel method to detect seven microcystins in hard clam and corbicula fluminea by liquid chromatography-tandem mass spectrometry. Author: Yang B, Xu JZ, Ding T, Wu B, Jing S, Ding SJ, Chen HL, Sheng CY, Jiang Y. Journal: J Chromatogr B Analyt Technol Biomed Life Sci; 2009 Nov 01; 877(29):3522-8. PubMed ID: 19748327. Abstract: A simple and reliable method to detect seven microcystins in hard clam and corbicula fluminea, based on liquid chromatography with electrospray ionization and tandem mass spectrometry (LC-ESI-MS/MS), was developed and validated. The sample preparation procedure includes extraction of tissue by methanol, followed by cleanup on a reversed-phase solid phase extraction (SPE) cartridge. With the optimized method, recoveries were between 43.7% and 92.3% for hard clam, 54.3% and 93.8% for corbicula fluminea, the relative standard deviations (RSD) were less than or equal to 16.2% and 15.7% in hard clam and corbicula fluminea at spiking levels of 1 microg/kg, 2 microg/kg and 5 microg/kg for MC-RR, MC-YR, MC-LR, and MC-LY, and 2 microg/kg, 5 microg/kg and 10 microg/kg for MC-LA, MC-LW and MC-LF, respectively, the limits of quantitation (LOQ) of this method were ranged from 0.7 microg/kg to 2.0 microg/kg.[Abstract] [Full Text] [Related] [New Search]