These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Development of a new high performance liquid chromatography method for measurement of coenzyme Q10 in healthy blood plasma.
    Author: Abdul-Rasheed OF, Farid YY.
    Journal: Saudi Med J; 2009 Sep; 30(9):1138-43. PubMed ID: 19750257.
    Abstract:
    OBJECTIVE: To improve a new high performance liquid chromatography (HPLC) method for coenzyme Q10 (CoQ10) measurement in human plasma. METHODS: We conducted this study in the Chemistry and Biochemistry Department, College of Medicine, Al-Nahrain University, Baghdad, Iraq from September 2004 to July 2005. Sixty healthy volunteers aged 28.51+/-7.30 years were participated in this study. All participants were non-smokers and under no any treatment for the last 3 days and fast for 12 hours before analysis. Coenzyme Q10 levels were measured by improved isocratic reversed phase HPLC technique after one-step ice-cold 1-propanol extraction step. RESULTS: The ultra violet (UV) detection, developed HPLC method gives a good linearity range between 0.1-4.0 mg/L. The coefficients of variations (CVs%) of within-day precision for CoQ10 in human plasma were 0.2-3.90%. Analytical recoveries were 95.5-101.30%, whereas the average CV% of CoQ10 reproducibility was 0.959%. CONCLUSION: A new improved rapid, precise, and sensitive HPLC method for the quantitation of CoQ10 is suitable for analysis of blood plasma samples.
    [Abstract] [Full Text] [Related] [New Search]