These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: [Long-term histological outcome for the larynx after Gore-Tex medialization thyroplasty]. Author: Sasai H, Watanabe Y, Miyahara H. Journal: Nihon Jibiinkoka Gakkai Kaiho; 2009 Aug; 112(8):623-7. PubMed ID: 19764545. Abstract: OBJECTIVES: We studied the long-term histological response of the human larynx to Gore-Tex implant and evaluated whether its biocompatibility. STUDY DESIGN: Retrospective histological study. METHODS: We conducted medialization laryngoplasty with expanded polytetrafluoroethylene (ePTEE or Gore-Tex) in a patient with vocal-fold paralysis. A strip of Gore-Tex was folded several times and inserted into the pocket preserving the inner perichondrium through a thyroid cartilage window. The patient died 12 months later and we histologically analyzed the larynx. RESULTS: No evidence was seen of foreign body granuloma, migration, extrusion, or infection. A modest inflammatory response with a fibrous capsule was present around implants. We confirmed the infiltration of abundant collagen and numerous fibroblasts into the microporous implant structure. CONCLUSION: Histological examination results suggest that Gore-Tex implants are safe and biocompatible with the human larynx. A slight inflammatory response and infiltration of fibrous tissue into the implant itself demonstrated that Gore-Tex implants are immunologically well tolerated.[Abstract] [Full Text] [Related] [New Search]