Search MEDLINE/PubMed
Title: Inappropriate ICD discharge due to T-wave oversensing in a patient with short QT syndrome. Author: Sun Y, Zhang P, Li X, Guo J. Journal: Pacing Clin Electrophysiol; 2010 Jan; 33(1):113-6. PubMed ID: 19793372. Abstract: A 45-year-old man with diagnosis of short QT syndrome underwent successful implantation of dual-chamber implantable cardioverter-defibrillator (ICD) to prevent sudden cardiac death. Ten days after implantation, the patient was treated by inappropriate ICD discharges for frequent T-wave oversensing. After careful reprogramming of the ICD, the T-wave oversensing was eliminated and no T-wave oversensing or inappropriate discharge was documented during 6-month follow-up period.[Abstract] [Full Text] [Related] [New Search]