These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Risk of failure of transvenous implantable cardioverter-defibrillator leads.
    Author: Borleffs CJ, van Erven L, van Bommel RJ, van der Velde ET, van der Wall EE, Bax JJ, Rosendaal FR, Schalij MJ.
    Journal: Circ Arrhythm Electrophysiol; 2009 Aug; 2(4):411-6. PubMed ID: 19808497.
    Abstract:
    BACKGROUND: Despite the positive effect on mortality in selected patients, implantable cardioverter-defibrillator therapy is also associated with potential malfunction of the implanted system. The present study provides the long-term lead failure rate in a large single-center cohort. METHODS AND RESULTS: Since 1992, a total of 2068 implantable cardioverter-defibrillator patients with 2161 defibrillation leads were prospectively collected. Data of the implant procedure and all follow-up visits were recorded. All cases of lead removal or capping or placing of an additional pace or sense lead were noted and analyzed. Lead models were grouped by manufacturer and approximate lead diameter in French. During a mean follow-up of 36 months, 82 (3.8%) cases of lead failure were identified. Cumulative incidence of lead failure at 1 year was 0.6%; at 5 years, 6.5%; and at 10 years, 16.4%. The highest risk of lead failure was found in small-diameter leads. Adjusted hazard ratio was 6.4 (95% CI, 3.2 to 12.8) for Medtronic 7F leads, when compared with all other leads. CONCLUSIONS: In this large single-center experience, the overall incidence of lead failure was 1.3 (95% CI, 1.0 to 1.6) per 100 lead-years. Comparison of different groups of leads shows major differences in event rates. Specific manufacturer's small-diameter defibrillation leads may have a higher risk of early lead failure.
    [Abstract] [Full Text] [Related] [New Search]