These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: High-performance liquid chromatography using UV detection for the simultaneous quantification of ropivacaine and bupivacaine in human plasma. Author: Gaudreault F, Drolet P, Varin F. Journal: Ther Drug Monit; 2009 Dec; 31(6):753-7. PubMed ID: 19809388. Abstract: A specific high-performance liquid chromatography assay coupled with UV detection has been developed and validated for the simultaneous determination of ropivacaine and bupivacaine in human plasma. A liquid-liquid back extraction procedure was used to increase specificity, and very good and consistent recoveries were obtained: 93%-95% for ropivacaine and 90%-96% for bupivacaine. The lowest limit of quantification was 4 and 8 ng/mL for ropivacaine and bupivacaine, respectively. The method was sensitive, reproducible (coefficient of variation <or= 6.9%), and accurate (96%-102%) and was applied successfully to study the pharmacokinetics of ropivacaine and bupivacaine in a perioperative context.[Abstract] [Full Text] [Related] [New Search]