  • Title: High-performance liquid chromatography using UV detection for the simultaneous quantification of ropivacaine and bupivacaine in human plasma.
    Author: Gaudreault F, Drolet P, Varin F.
    Journal: Ther Drug Monit; 2009 Dec; 31(6):753-7. PubMed ID: 19809388.
    Abstract:
    A specific high-performance liquid chromatography assay coupled with UV detection has been developed and validated for the simultaneous determination of ropivacaine and bupivacaine in human plasma. A liquid-liquid back extraction procedure was used to increase specificity, and very good and consistent recoveries were obtained: 93%-95% for ropivacaine and 90%-96% for bupivacaine. The lowest limit of quantification was 4 and 8 ng/mL for ropivacaine and bupivacaine, respectively. The method was sensitive, reproducible (coefficient of variation <or= 6.9%), and accurate (96%-102%) and was applied successfully to study the pharmacokinetics of ropivacaine and bupivacaine in a perioperative context.