Title: Subserosal misplacement of Essure device manifested by late-onset acute pelvic pain. Author: Mahmoud MS, Fridman D, Merhi ZO. Journal: Fertil Steril; 2009 Dec; 92(6):2038.e1-3. PubMed ID: 19819445. Abstract: OBJECTIVE: To increase awareness of the potential to present with late-onset acute pelvic pain secondary to subserosal misplacement of an Essure device (Conceptus Inc., Mountain View, CA). DESIGN: Case report. SETTING: University-affiliated teaching hospital. PATIENT(S): A 30-year-old woman who was seen with severe left lower quadrant pain 4 months after elective sterilization with an Essure device placed under local anesthesia. Mild to moderate resistance was encountered in the placement of the device in the left fallopian tube. INTERVENTION(S): Hysterosalpingogram showing patency of the left fallopian tube and operative laparoscopy. MAIN OUTCOME MEASURE(S): Laparoscopic removal of the Essure device with left salpingectomy. RESULT(S): The patient was free of pain and was discharged home the same day of the laparoscopic procedure. CONCLUSION(S): This report reinforces the need to consider a misplaced Essure device in the differential diagnosis of late-onset acute pelvic pain in women who had difficult placement of the device.[Abstract] [Full Text] [Related] [New Search]