  • Title: The 2.5 mm PushLock suture anchor system versus a traditional suture anchor for ulnar collateral ligament injuries of the thumb: a biomechanical study.
    Author: Jarrett CD, McGillivary GR, Hutton WC.
    Journal: J Hand Surg Eur Vol; 2010 Feb; 35(2):139-43. PubMed ID: 19828569.
    Abstract:
    We compared the biomechanical strength of the 2.5 mm PushLock suture anchor with a traditional Bio-SutureTak suture anchor in repair of ulnar collateral ligament injuries. Iatrogenic ulnar collateral ligament injuries in 18 cadaveric thumbs were repaired and used to test for load to failure and cyclic loading. The average force required to generate a 2 mm gap was 7.7 N for the 2.5 mm PushLock and 6.3 N for the Bio-SutureTak (p = 0.04). The ultimate load to failure was 28.0 N for the 2.5 mm PushLock and 18.8 N for the Bio-SutureTak (p = 0.16). There were no statistical differences between the two suture anchors under cyclic loading. The 2.5 mm PushLock suture anchor provides significantly stronger resistance to 2 mm gap formation at the repair site and is less likely to fail at the suture-ligament interface. However, there was no difference in the load to failure between the two suture anchors.
    [Abstract] [Full Text] [Related] [New Search]