These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Backscattered radiation into a transmission ionization chamber: measurement and Monte Carlo simulation.
    Author: Yoshizumi MT, Yoriyaz H, Caldas LV.
    Journal: Appl Radiat Isot; 2010; 68(4-5):586-8. PubMed ID: 19884019.
    Abstract:
    Backscattered radiation (BSR) from field-defining collimators can affect the response of a monitor chamber in X-radiation fields. This contribution must be considered since this kind of chamber is used to monitor the equipment response. In this work, the dependence of a transmission ionization chamber response on the aperture diameter of the collimators was studied experimentally and using a Monte Carlo (MC) technique. According to the results, the BSR increases the chamber response of over 4.0% in the case of a totally closed collimator and 50 kV energy beam, using both techniques. The results from Monte Carlo simulation confirm the validity of the simulated geometry.
    [Abstract] [Full Text] [Related] [New Search]