  • Title: T wave oversensing and low percentage of biventricular pacing in cardiac resynchronization therapy.
    Author: Lezcano AO, Aracama JM, Ayestaran VU, Urra FG.
    Journal: Cardiol J; 2009; 16(6):580-1. PubMed ID: 19950098.
    Abstract:
    The success of cardiac resynchronization therapy (CRT) depends on a high percentage of ventricular pacing. We present the electrocardiography and electrograms of a patient who underwent an implantable cardioverter-defibrillator (ICD) with CRT implantation showing a low percentage of ventricular pacing as a result of T wave oversensing of paced QRS. The patient showed no clinical improvement. We suggest finding a good sensitivity of F waves in induced ventricular fibrillation during ICD implantation in order to overcome the harm of possible oversensing.
    [Abstract] [Full Text] [Related] [New Search]