These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Hydrodynamics of aortic cannulae during extracorporeal circulation in a mock aortic arch aneurysm model. Author: Minakawa M, Fukuda I, Igarashi T, Fukui K, Yanaoka H, Inamura T. Journal: Artif Organs; 2010 Feb; 34(2):105-12. PubMed ID: 19995359. Abstract: This study was designed to analyze flow pattern, velocity, and strain on the aortic wall of a glass aortic arch aneurysm model during the extracorporeal circulation, and to elucidate the characteristics of flow pattern in three different aortic cannulae. Different patterns of large vortices and helical flow were made by each cannula. With the curved end-hole cannula, the high velocity flow (approximately 0.6-0.8 m/s) was blowing to the aneurismal wall without attenuating the strain rate tensor (approximately 0.2-0.25/s). With the dispersion cannula and the Soft-Flow cannula, cannular jet was attenuated in the ascending aorta creating a large vortex at a velocity less than 0.5 m/s, and the strain rate tensor on the aneurismal wall was small (less than 0.15/s). In conclusion, end-hole cannula should not be used in the operation of aortic arch aneurysm. Dispersion-type aortic cannulae were less invasive on the aortic arch aneurismal wall, but particular attention to alternative cannulation sites should be paid in cases with severe atherosclerosis on the ascending aortic wall.[Abstract] [Full Text] [Related] [New Search]