  • Title: Diffractive-refractive optics: low-aberration Bragg-case focusing by precise parabolic surfaces.
    Author: Oberta P, Mikulík P, Kittler M, Hrdý J, Peverini L.
    Journal: J Synchrotron Radiat; 2010 Jan; 17(1):36-40. PubMed ID: 20029109.
    Abstract:
    Based on analytical formulae calculations and ray-tracing simulations a low-aberration focal spot with a high demagnification ratio was predicted for a diffractive-refractive crystal optics device with parabolic surfaces. Two Si(111) crystals with two precise parabolic-shaped grooves have been prepared and arranged in a dispersive position (+,-,-,+) with high asymmetry. Experimental testing of the device at beamline BM05 at the ESRF provided a focal spot size of 38.25 microm at a focal distance of 1.4 m for 7.31 keV. This is the first experiment with a parabolic-shaped groove; all previous experiments were performed with circular grooves which introduced extreme aberration broadening of the focal spot. The calculated and simulated focal size was 10.8 microm at a distance of 1.1 m at 7.31 keV. It is assumed that the difference between the measured and calculated/simulated focal spot size and focal distance is due to insufficient surface quality and to alignment imperfection.