These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: A simple approximate formula for the physical focal length of spherically focused transducers.
    Author: Huang JH, Ding D.
    Journal: IEEE Trans Ultrason Ferroelectr Freq Control; 2009 Dec; 56(12):2764-8. PubMed ID: 20040414.
    Abstract:
    A simple approximate formula is presented to describe the relationship between the physical focal lengths and the geometric focal lengths of spherically focused transducers with ultrasonic fields. This simple expression applies regardless of whether the focus strength is strong or weak, and regardless of whether the fluids are non-attenuating or attenuating. Analysis results show that the average error of the simple approximate formula is certainly less than 2% relative to the numerical solution.
    [Abstract] [Full Text] [Related] [New Search]