These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Twiddling with an implantable cardioverter-defibrillator. Author: McCann CJ, Beelen R, Geelen P. Journal: Acta Cardiol; 2009 Oct; 64(5):695. PubMed ID: 20058522. Abstract: A 64-year-old woman underwent implantation of a single-chamber implantable cardioverter-defibrillator. The active fixation lead was positioned in the right ventricular apex and the generator in a left pre-pectoral pocket. Device interrogation 4weeks after implantation revealed inappropriate sensing and failure of pacing. A chest X-ray showed the entire lead coiled behind the device. The patient admitted manipulating the device and a diagnosis of Twiddler's syndrome was made. After repositioning the lead, the generator was re-located in the sub-pectoral position to help prevent recurrence by reducing the ability to externally manipulate the device.[Abstract] [Full Text] [Related] [New Search]