Search MEDLINE/PubMed
Title: Amperometric catechol biosensor based on polyaniline-polyphenol oxidase. Author: Tan Y, Guo X, Zhang J, Kan J. Journal: Biosens Bioelectron; 2010 Mar 15; 25(7):1681-7. PubMed ID: 20060283. Abstract: A novel catechol biosensor was described based on the immobilization of polyphenol oxidase (PPO) into polyaniline (PANI), which was easily constructed by direct electropolymerization of aniline in a solution containing ionic liquid, 1-ethyl-3-methylimidazolium ethyl sulfate (EMIES). The developed biosensor for the detection of catechol has a linear range of 1.25-150 micromol dm(-3). The maximum response current (I(max)) and the Michaelis-Menten constant (k'(m)) are 0.62 microA and 146 micromol dm(-3), respectively. The activation energy (E(a)) of the PPO catalytic reaction is 31.1 kJ mol(-1) in the B-R buffer. The biosensor shows good reproducibility (a relative standard deviation of 3.1% was obtained) and remarkable long-term stability (it retains 75% of the original activity after four months). The effects of potential and pH on the response current of the biosensor are also described.[Abstract] [Full Text] [Related] [New Search]