These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Standardisation of neurophysiological norm values. Relevance of the position of the heating element and the temperature measurement sensor.
    Author: Husstedt IW, Grotemeyer KH, Schlake HP.
    Journal: Electromyogr Clin Neurophysiol; 1991; 31(1):61-4. PubMed ID: 2009828.
    Abstract:
    At a "constant" temperature of 37 degrees C, we measured the influence of the position of the temperature measurement sensor and the infrared heating element on the nervous conduction velocity (NCV) and the delay in latency after paired stimulation (LPSS) of the sural nerve. Temperature differences of up to 6.2 degrees C resulted from alteration of four cm in position of the temperature measurement sensor and the infrared heating element. NCV in the test subjects changed from 54.2 + -4.5 to 57.8 + -4.0 m/sec and the LPSS from 4.0 + -2.2% to 2.0 + -1.0%. The patients showed an increase of NCV from 54.1 + -5.0 to 56.8 + -5.7 m/sec and a decrease of LPSS from 9.9 + -5.7% to 6.5 + -3.9%. Our investigation shows that even the position of the temperature measurement sensor and the infrared heating element form a very appreciable component of the standardization method and decide on normal values. The consistent implementation of these results for routine purposes restricts the range of error of neurophysiological measurements, as well as improving the reproducibility and relevance of borderline findings.
    [Abstract] [Full Text] [Related] [New Search]