These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: [Refraction by atrophine cycloplegia]. Author: Hunold W, Auffarth G. Journal: Ophthalmologica; 1991; 202(1):26-32. PubMed ID: 2017328. Abstract: A new method for refractive measurements with atropine cycloplegia was tested in 90 children with squint. Measurements were taken by an autorefractor (Canon R 10) 90 min after application of 2 atropine eye drops and after 3 days of receiving 1 atropine eye drop 3 times daily, i.e. 9 drops per eye. Results were also compared to conventional "objective" refractometry by means of a hand refractometer (Rodenstock PR 50). According to this method, the total refraction, spherical and cylindric power and the cylinder axis are in agreement with refractive data in 80-90% of the children after 3-day atropinization. The additional cycloplegic effect of atropine given for 3 days is only 0.5 dptr compared to the application of 2 drops on the first day. So for practical purposes the refractive data evaluated after 90-min atropinization can be used as a basis for the prescription of glasses.[Abstract] [Full Text] [Related] [New Search]