  • Title: Multicenter evaluation of the first automated Elecsys sFlt-1 and PlGF assays in normal pregnancies and preeclampsia.
    Author: Schiettecatte J, Russcher H, Anckaert E, Mees M, Leeser B, Tirelli AS, Fiedler GM, Luthe H, Denk B, Smitz J.
    Journal: Clin Biochem; 2010 Jun; 43(9):768-70. PubMed ID: 20206155.
    Abstract:
    OBJECTIVES: Performance evaluation of Elecsys sFlt-1 and PlGF assays. DESIGN AND METHODS: Within-, between-run, total imprecision, functional sensitivity, inter-laboratory comparison, method comparison and lot-to-lot reproducibility were evaluated. RESULTS: Within- and between-run CVs were below 4% for sFlt-1 >60 and PlGF > 20 pg/mL. Total imprecision CVs were below 4.3%. Functional sensitivity was < 5 pg/mL. Inter-laboratory CVs were <5%. Elecsys correlated well with Quantikine VEGF-R1 (r=0.960) and PlGF (r=0.968). Lot-to-lot comparisons yielded highly correlated results (r>0.999). In healthy pregnancies, the median levels of sFlt-1 remained constant in first (1107 pg/mL) and second trimesters (1437 pg/mL) but increased in the third trimester (2395 pg/mL), while median PlGF levels increased in the first (30 pg/mL) and second trimesters (279 pg/mL) and peaked at 29 to 32 weeks (626 pg/mL) and decreased thereafter (340 pg/mL). The sFlt-1/PlGF ratio is highest in the first trimester (median: 28) but remained constant in the second (median: 4.7) and third trimesters (median: 5.1). In PE/HELPP samples matched for gestational age the sFlt-1 levels were significantly higher (6894-34,624 pg/mL), whereas PlGF levels were lower (9.2-80 pg/mL) and the median sFlt-1/PlGF ratio is much higher (461; range: 121-2614) than in apparently healthy pregnancies (3.6; range: 0.3-105). CONCLUSION: The new Roche Elecsys sFlt-1 and PlGF immunoassay showed excellent precision and reliability. There was a clear difference in the Elecsys sFlt-1/PlGF ratio between samples obtained from women with apparently normal pregnancy at the time of blood collection and those diagnosed with PE/HELLP at the same age of gestation.
    [Abstract] [Full Text] [Related] [New Search]