These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Accuracy and precision of fourteen pulse oximeters. Author: Hannhart B, Haberer JP, Saunier C, Laxenaire MC. Journal: Eur Respir J; 1991 Jan; 4(1):115-9. PubMed ID: 2026231. Abstract: Two sets of seven pulse oximeters (Criticare CSI-502; Nellcor-N200; Datex-Satlite; Physio-Control-Lifestat 1600; Critikon-Oxyshuttle; Ohmeda-Biox 3700; Ohmeda-Biox 3740; Radiometer-Oxi; Spectramed-Pulsat; Kontron-7840; Biochem-Ox2000; Invivo-4500; Engström-EOS; Novametrix-505) were studied in two groups of eight healthy subjects, aged 26-50 yrs. The transcutaneous oxygen saturation (SpO2) was compared with arterial oxygen saturation (SaO2) measured in simultaneously with drawn blood samples (OSM2 Radiometer) at four 20 min steady-state levels of inspired oxygen fraction (FIO2) (0.21, 0.10, 0.08 and 0.07; SaO2 99-55%) in a conditioned chamber. Both the error in accuracy (mean SpO2-SaO2 difference), and the error in precision (SD of the differences) remained less than 3% for the two highest FIO2 levels (SaO2 greater than 83%) but, during deeper hypoxia, they were increased to 8% and 5%, respectively. An instrumental systematic bias affected accuracy in particular. We concluded that a good agreement between SpO2 and SaO2, as reflected by the Bartko's intraclass coefficient, was observed in nine instruments.[Abstract] [Full Text] [Related] [New Search]