These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: High-throughput fluorescence detection using an integrated zone-plate array.
    Author: Schonbrun E, Abate AR, Steinvurzel PE, Weitz DA, Crozier KB.
    Journal: Lab Chip; 2010 Apr 07; 10(7):852-6. PubMed ID: 20300671.
    Abstract:
    Microfluidic devices enable massive parallelization of sample manipulation and delivery, but a similarly parallelized and integrated optical detection system does not yet exist. Standard large numerical aperture wide field or scanning optical systems are not capable of the large field of view and detection sensitivity required to collect fluorescence from parallel arrays of microfluidic devices. Instead, we present a fluorescence measurement platform based on a microfabricated zone-plate array integrated into a parallelized microfluidic device. The zone-plate array is orientated so that a single high numerical aperture zone plate is aligned to read out the fluorescence from each of 64 output channels of a drop-making device. The parallelization of microfluidics and optics produces an integrated system capable of analysis of nearly 200,000 drops per second.
    [Abstract] [Full Text] [Related] [New Search]