These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Catheter ablation of accessory atrioventricular pathways (Wolff-Parkinson-White syndrome) by radiofrequency current.
    Author: Jackman WM, Wang XZ, Friday KJ, Roman CA, Moulton KP, Beckman KJ, McClelland JH, Twidale N, Hazlitt HA, Prior MI.
    Journal: N Engl J Med; 1991 Jun 06; 324(23):1605-11. PubMed ID: 2030716.
    Abstract:
    BACKGROUND: Surgical or catheter ablation of accessory pathways by means of high-energy shocks serves as definitive therapy for patients with Wolff-Parkinson-White syndrome but has substantial associated morbidity and mortality. Radiofrequency current, an alternative energy source for ablation, produces smaller lesions without adverse effects remote from the site where current is delivered. We conducted this study to develop catheter techniques for delivering radiofrequency current to reduce morbidity and mortality associated with accessory-pathway ablation. METHODS: Radiofrequency current (mean power, 30.9 +/- 5.3 W) was applied through a catheter electrode positioned against the mitral or tricuspid annulus or a branch of the coronary sinus; when possible, delivery was guided by catheter recordings of accessory-pathway activation. Ablation was attempted in 166 patients with 177 accessory pathways (106 pathways in the left free wall, 13 in the anteroseptal region, 43 in the posteroseptal region, and 15 in the right free wall). RESULTS: Accessory-pathway conduction was eliminated in 164 of 166 patients (99 percent) by a median of three applications of radiofrequency current. During a mean follow-up (+/- SD) of 8.0 +/- 5.4 months, preexcitation or atrioventricular reentrant tachycardia returned in 15 patients (9 percent). All underwent a second, successful ablation. Electrophysiologic study 3.1 +/- 1.9 months after ablation in 75 patients verified the absence of accessory-pathway conduction in all. Complications of radiofrequency-current application occurred in three patients (1.8 percent): atrioventricular block (one patient), pericarditis (one), and cardiac tamponade (one) after radiofrequency current was applied in a small branch of the coronary sinus. CONCLUSIONS: Radiofrequency current is highly effective in ablating accessory pathways, with low morbidity and no mortality.
    [Abstract] [Full Text] [Related] [New Search]