These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: A highly sensitive LC-MS/MS method for the determination of S-citalopram in rat plasma: application to a pharmacokinetic study in rats.
    Author: Suresh PS, Giri S, Husain R, Mullangi R.
    Journal: Biomed Chromatogr; 2010 Oct; 24(10):1052-8. PubMed ID: 20352615.
    Abstract:
    A highly sensitive, rapid assay method has been developed and validated for the estimation of S-citalopram (S-CPM) in rat plasma with liquid chromatography coupled to tandem mass spectrometry with electrospray ionization in the positive-ion mode. The assay procedure involves a simple liquid-liquid extraction of S-CPM and phenacetin (internal standard, IS) from rat plasma with t-butyl methyl ether. Chromatographic separation was operated with 0.2% formic acid:acetonitrile (20:80, v/v) at a flow rate of 0.50 mL/min on a Symmetry Shield RP(18) column with a total run time of 3.0 min. The MS/MS ion transitions monitored were 325.26 → 109.10 for S-CPM and 180.10 → 110.10 for IS. Method validation and pre-clinical sample analysis were performed as per FDA guidelines and the results met the acceptance criteria. The lower limit of quantitation achieved was 0.5 ng/mL and the linearity was observed from 0.5 to 5000 ng/mL. The intra- and inter-day precisions were in the range of 1.14-5.56 and 0.25-12.3%, respectively. This novel method has been applied to a pharmacokinetic study and to estimate brain-to-plasma ratio of S-CPM in rats.
    [Abstract] [Full Text] [Related] [New Search]