  • Title: Continuous on-line monitoring of left ventricular function with a new nonimaging detector:validation and clinical use in the evaluation of patients post angioplasty.
    Author: Breisblatt WM, Schulman DS, Follansbee WP.
    Journal: Am Heart J; 1991 Jun; 121(6 Pt 1):1609-17. PubMed ID: 2035375.
    Abstract:
    A new miniaturized nonimaging radionuclide detector (Cardioscint, Oxford, England) was evaluated for the continuous on-line assessment of left ventricular function. This cesium iodide probe can be placed on the patient's chest and can be interfaced to an IBM compatible personal computer conveniently placed at the patient's bedside. This system can provide a beat-to-beat or gated determination of left ventricular ejection fraction and ST segment analysis. In 28 patients this miniaturized probe was correlated against a high resolution gamma camera study. Over a wide range of ejection fraction (31% to 76%) in patients with and without regional wall motion abnormalities, the correlation between the Cardioscint detector and the gamma camera was excellent (r = 0.94, SEE +/- 2.1). This detector system has high temporal (10 msec) resolution, and comparison of peak filling rate (PFR) and time to peak filling (TPFR) also showed close agreement with the gamma camera (PFR, r = 0.94, SEE +/- 0.17; TPFR, r = 0.92, SEE +/- 6.8). In 18 patients on bed rest the long-term stability of this system for measuring ejection fraction and ST segments was verified. During the monitoring period (108 +/- 28 minutes) only minor changes in ejection fraction occurred (coefficient of variation 0.035 +/- 0.016) and ST segment analysis showed no significant change from baseline. To determine whether continuous on-line measurement of ejection fraction would be useful after coronary angioplasty, 12 patients who had undergone a successful procedure were evaluated for 280 +/- 35 minutes with the Cardioscint system.(ABSTRACT TRUNCATED AT 250 WORDS)
    [Abstract] [Full Text] [Related] [New Search]