These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Kinetic analysis and fragment screening with Fujifilm AP-3000.
    Author: Rich RL, Myszka DG.
    Journal: Anal Biochem; 2010 Jul 15; 402(2):170-8. PubMed ID: 20371220.
    Abstract:
    We evaluated the performance of Fujifilm's new AP-3000 surface plasmon resonance biosensor for kinetic analysis and fragment screening. Using carbonic anhydrase II as a model system, we characterized a set of 10 sulfonamide-based inhibitors that range in molecular mass from 98 to 341Da and approximately 10,000-fold in affinity (0.4mM to 20nM). Although the data collected from the AP-3000 were generally similar to those collected using a Biacore T100, the AP-3000's stop-flow analyte delivery system complicated the shapes of the association- and dissociation-phase binding responses. We illustrate how reasonable estimates of the kinetic rate constants can be extracted from AP-3000 data by limiting data analysis to only the regions of the responses collected during flow conditions. We also provide an example of the results obtained for a fragment-screening study with the AP-3000, which is the ideal application of this technology.
    [Abstract] [Full Text] [Related] [New Search]