Title: A comparative study on the accuracy of noninvasive thermometers. Author: Fadzil FM, Choon D, Arumugam K. Journal: Aust Fam Physician; 2010 Apr; 39(4):237-9. PubMed ID: 20372685. Abstract: BACKGROUND: This study assessed the concordance of the temperatures of the digital, liquid crystal forehead and digital infrared tympanic thermometers with the mercury in glass thermometer. METHODS: Temperatures in degrees celsius were taken simultaneously using the four thermometers in 207 patients at the casualty department of a Malaysian hospital. The Bland Altman statistical test was used to assess the concordance by the 95% limits of agreement between the three newer thermometers and the mercury in glass thermometer. RESULTS: The digital thermometer gave the best concordance (limits of agreement 0.48-0.59 degrees C). The liquid crystal forehead thermometer gave the least concordance (limits of agreement -1.14-0.98 degrees C). The digital infrared tympanic was in between (limits of agreement -0.88-0.85 degrees C). DISCUSSION: The digital thermometer provides the best agreement with the mercury in glass thermometer. The infrared tympanic thermometer may be a preferable option for the uncooperative patient. The liquid crystal forehead thermometer is best used at home.[Abstract] [Full Text] [Related] [New Search]