These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Detection of prostate-specific antigen with a paired surface plasma wave biosensor. Author: Su LC, Chen RC, Li YC, Chang YF, Lee YJ, Lee CC, Chou C. Journal: Anal Chem; 2010 May 01; 82(9):3714-8. PubMed ID: 20373787. Abstract: In this study, we demonstrated that an amplitude-sensitive paired surface plasma wave biosensor (PSPWB) is capable of real-time detection of prostate-specific antigen (PSA) in diluted human serum without labeling. Experimentally, the detection limit of PSPWB was 8.4 x 10(-9) refractive index unit (RIU) and the PSPWB could measure PSA in a phosphate buffered saline solution from 10 fg/mL ( approximately 300 aM) to 100 pg/mL ( approximately 3 pM) successfully, with demonstration of a linear relationship between PSA concentrations and surface plasmon resonance (SPR) signals. Therefore, results were obtained over a wide dynamic range 5 orders of magnitude for analyte concentration. In addition, the PSPWB successfully detected PSA in diluted human serum as well. These experimental results indicate that the PSPWB is capable of detection with high sensitivity over a wide range by using SPR-based biosensors and has a capability of detecting biological analytes in clinical sample without complicated operating procedures.[Abstract] [Full Text] [Related] [New Search]