These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Stent fractures in the Hemobahn/Viabahn stent graft after endovascular popliteal aneurysm repair.
    Author: Tielliu IF, Zeebregts CJ, Vourliotakis G, Bekkema F, van den Dungen JJ, Prins TR, Verhoeven EL.
    Journal: J Vasc Surg; 2010 Jun; 51(6):1413-8. PubMed ID: 20382495.
    Abstract:
    OBJECTIVE: During the last decade, endovascular repair of popliteal artery aneurysms (PAAs) has become a valid alternative to open repair. This study analyzes the incidence and origin of stent graft fractures after endovascular repair, its impact on patency, and strategies to prevent fractures. METHODS: Data of 78 atherosclerotic PAAs in 64 patients were gathered in a prospectively-held database from 1998 to 2009. All x-rays were reviewed to detect stent fractures. Only circumferential fractures were included for analysis; localized strut fractures were excluded. Clinical endpoints were circumferential stent fracture, occlusion, and clinical status of the patient. RESULTS: Mean follow-up time was 50 months (range, 1-127 months). Fifteen circumferential stent fractures occurred in 13 (16.7%) patients. The majority of stent fractures (93.3%) were associated with the use of multiple stent grafts. At univariate analysis, younger age was identified as the only significant predictor for stent fracture (P = .007). The cumulative stent fracture-free survival was estimated at 78% and 73% at 5- and 10-year follow-up, respectively. The cumulative primary patency rate, defined as time to occlusion, was not different for the fracture group compared with the nonfracture group (P = .284). CONCLUSIONS: The incidence of stent fractures after endovascular PAA repair is probably underreported in the literature. Stent graft fractures mainly occur at overlap zones and are associated with younger age of the patient. Fracture of the stent did not significantly influence patency of the stent graft.
    [Abstract] [Full Text] [Related] [New Search]