These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Experimental and Monte Carlo determination of the TG-43 dosimetric parameters for the model 9011 THINSeed brachytherapy source.
    Author: Kennedy RM, Davis SD, Micka JA, DeWerd LA.
    Journal: Med Phys; 2010 Apr; 37(4):1681-8. PubMed ID: 20443489.
    Abstract:
    PURPOSE: AAPM TG-43 brachytherapy dosimetry parameters for a new, smaller diameter 1251 brachytherapy source (THINSeed, model 9011) were determined using LiF:Mg,Ti thermoluminescent dosimeter (TLD-100) microcubes and Monte Carlo simulations. METHODS: Two polymethyl methacrylate phantoms were machined to hold TLD-100 microcubes at specific locations for the experimental determination of the radial dose function, dose-rate constant, and anisotropy functions of the new source. The TG-43 parameters were also calculated using Monte Carlo simulations. For comparison, the model 6711 source was also investigated. RESULTS: Experimental results for both models 9011 and 6711 sources showed good agreement with Monte Carlo values, as well as with previously published values. CONCLUSIONS: The TG-43 parameters for the new source model are similar to those of model 6711; however, they represent two separate sources and TG-43 parameters used in treatment planning must be source specific.
    [Abstract] [Full Text] [Related] [New Search]