These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: [Feedback control of depth of anesthesia during propofol administration. Bispectral index as the controlled variable]. Author: Janda M, Bajorat J, Simanski O, Kähler R, Pohl B, Nöldge-Schomburg GF, Hofmockel R. Journal: Anaesthesist; 2010 Jul; 59(7):621-7. PubMed ID: 20461346. Abstract: BACKGROUND: The objective of this study was to evaluate the performance of a new system for closed-loop control of propofol administration using the bispectral index (BIS) under total intravenous anesthesia in the index values of middle-to-deep depth of anaesthesia. METHODS: In this study 20 adult patients anesthetized with propofol and remifentanil were investigated. The propofol infusion was carried out using a fuzzy-PD+I controller with a target BIS value of 40. RESULTS: Closed-loop control was able to provide maintenance of anesthesia and adequate operating conditions for all patients. The following quality control criteria were calculated: median performance error (MDPE; 0.16%, SD +/-1.4%), median absolute performance error (MDAPE; 6.9%, SD +/-2.8%) and wobble (6.8%, SD +/-2.5%). CONCLUSION: The present study showed the clinical feasibility of the controller compared to existing devices regarding a high level of quality criteria of a model with an implemented fuzzy-PD+I structure controlling depth of hypnosis.[Abstract] [Full Text] [Related] [New Search]