These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Evaluation of H*(10) using the developed spherical type neutron dose monitor.
    Author: Bhuiya SH, Yamanishi H, Uda T.
    Journal: Radiat Prot Dosimetry; 2010 Oct; 141(3):217-21. PubMed ID: 20573684.
    Abstract:
    An instrument for evaluating the neutron ambient dose equivalent has been developed. It has the characteristic of uniform response to wide energy of neutrons. The monitor is four-layered spherically shaped, based on moderation and absorption of neutrons. Neutron dose can be evaluated from the linear combination of three specific responses of thermoluminescent dosimeters (TLDs), which are located at three depths in the moderator. TLDs were arranged between layers of two consecutive depths on 12 radial axes at even intervals so that the monitor is equally sensitive to all directions of neutrons. In order to verify the usefulness of dose evaluation by the monitor, irradiation experiments were conducted at the FRS, JAEA. The D2O-moderated 252Cf was used for the calibration of the monitor. Experiments were also conducted by using two neutron sources of 252Cf bare and 241Am-Be. As a result, the evaluated dose for each irradiation was obtained close to the actual irradiated dose. It was confirmed that the method of dose evaluation by the developed monitor can be applied to practical neutron fields where the distance of neutron source is unknown.
    [Abstract] [Full Text] [Related] [New Search]