These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Low-frequency RF hyperthermia: IV--A 27 MHz hybrid applicator for localized deep tumor heating. Author: Franconi C, Raganella L, Tiberio CA. Journal: IEEE Trans Biomed Eng; 1991 Mar; 38(3):287-93. PubMed ID: 2066143. Abstract: A 27 MHz dual-device applicator of novel design, which is aimed to heat noninvasively with improved safety tumor masses at depth, is proposed. A substantially localized temperature gain is obtained by superimposing two delocalized low RF frequency and phase-coherent current distributions, which are launched to constructively interfere over a limited region emcompassing the tumor volume. An hybrid applicator (HA) has been developed, integrated one capacitive and one inductive heating device, and has been assessed on a 20 cm diameter cylindrical fat-muscle phantom. The interference pattern is characterized by a deep broad SAR maximum and by the disappearance of the central null SAR value typical of single inductive devices. An 80% SAR useful therapeutic volume (UTV) of a near-cylindrical shape of about 800 cm3 is obtained with a penetration of 6-8 cm for the phantom surface, with a noncritical axial length of approximately 21 cm. The UTV may be somehow controlled in size and penetration. These results are obtained with the tissue-like medium surrounding the UTV heated uniformly and safely to a temperature pedestal below the therapeutic temperature with about half RF power values to each of the heating devices.[Abstract] [Full Text] [Related] [New Search]